Inspection and Care Instruments before Sterilising

All instruments that have been through any cleaning procedure should be inspected to ensure they are clean, functional and in good condition.

Any instruments that are blunt, bent or damaged or show any signs of pitting or other corrosion should be discarded. An illuminated magnifier is recommended because it makes it much easier to see residual contamination, debris or damage.

Staff should ensure that: there is free movement of all parts and that joints do not stick; the edges of clamping instruments meet with no overlap and that teeth mesh together; scissor edges meet to the tip and move freely across each other with no overlap or burrs (rough edges); all screws on jointed instruments are tight and have not become loose during use:

Instruments should be inspected for any visible soiling, such as blood. It is especially important to check joints, hinges or the serrated surfaces of jaws, which are difficult to clean. If there is any residual contamination, the instrument should be rejected and should undergo another cycle of the cleaning process.

Occasional use of a lubricant may be required where hinged instruments are found to be stiff. A non-oil-based lubricant should be used to avoid it interfering (that is, preventing the steam coming into contact with the instrument surface) with the sterilisation process.

Instruments may become damaged during use or suffer from general wear and tear over their life span. If devices are found to be faulty or damaged during inspection and function-testing, or if users identify that they are faulty, they should be taken out of use and either repaired or replaced. Instruments for repair should be decontaminated, labelled to identify they have been through the decontamination process, and then returned to either the manufacturer or a reputable repair company.

Hand pieces should be lubricated according to the manufacturer’s instructions. Those that have been processed in a washer-disinfector might have had the lubricant removed and require lubrication again before going into the steriliser.

A separate canister of lubricant should be used for cleaned instruments. The canisters should be labelled so that it is clear which canister is used for unclean instruments, and which is used for instruments that have been cleaned in a washer-disinfector. Another canister, for use with hand pieces after sterilisation, might be required if the manufacturer recommends it.

Inadequate lubrication can lead to unnecessary damage to the internal mechanism.