Instrument storage and wrapping
- Wrapped instruments may be stored up to 1 year.
- Unwrapped instruments in the clinical area – Max storage 1 day:
- Instruments should be dry; and protected from contamination, for example in mini-racks placed in cupboards or in covered drawer inserts.
- Instruments should not be stored on open work surfaces, particularly in clinical areas.
- It is important that practices have well developed protocols and procedures in place to prevent contamination of these instruments by ensuring that those required for a particular patient are removed from their protected environment before treatment commences.
- This eliminates the need to open cupboard doors or drawers during patient treatment. If an instrument does need to be retrieved from a cupboard or drawer during treatment, the practice should have protocols in place to prevent contamination and to ensure that staff hands are clean and that new gloves are donned before handling unwrapped sterilised instruments.
- Regard all instruments set out for each patient as contaminated after the treatment, whether or not they have been used.
- Instruments that are kept unwrapped should be reprocessed at the end of the working day, regardless of whether they have been used. Alternatively, instruments can be reprocessed at the beginning of the next working day.
- Unwrapped instruments in a non-clinical area – Max storage 1 week:
- Non-clinical area in this context is designated as a clinical area not in current use, or in a clean area of a separate decontamination room. Instruments should still be stored as follows:
- Dry and protected from contamination, for example in mini-racks placed in cupboards, or in covered drawer inserts. Instruments should not be placed on open work surfaces.
- Non-clinical area in this context is designated as a clinical area not in current use, or in a clean area of a separate decontamination room. Instruments should still be stored as follows: