Sterilisation And Storage
Saturated steam under pressure delivered at the highest temperature compatible with the product, is the preferred method for the sterilisation of most instruments used in the clinical setting.
To facilitate sterilisation, load items should first be thoroughly cleaned and disinfected ,In the case of newer machines, the parameters monitored for each cycle will be stored and/or available as a print-out to provide a short-term record. The use of automated data-loggers or interfaced small computer-based recording systems is acceptable, provided the records are kept securely and backed-up. These records should be copied, as the quality of the print-out fades over time. Manual recording using a logbook is also acceptable, and in any case will be a necessity if a machine does not have any automatic print-out function. The record should, at minimum, document the absence of a failure warning or the temperature/pressure achieved as appropriate to the indications provided. Records are required for every sterilisation cycle. It is recommended that records be maintained for not less than two years.
It is likely that steam sterilisers used in practices will have a chamber volume of less than 60 L and thus be considered to be small devices within the standards applied by national and international bodies.
Daily Testing and Housekeeping Tasks
Some sterilisers require a warm-up cycle before instruments can be processed. The manufacturer’s instruction manual should be consulted to find out whether this is the case.
The daily tests should be performed by the operator or user and will normally consist of:
- An automatic control test (all bench top sterilisers) in line with manufacturers’ instructions;
- A record of the temperature and pressure achieved at the daily test, to ensure this is satisfactory before the autoclave is used for sterilizing instruments.
These outcomes should be recorded in the logbook together with the date and signature of the operator.
The tests may be carried out at the same time. The manufacturer’s advice should be sought on whether the daily tests can be carried out while instruments are being reprocessed.
Before carrying out the daily tests, the user should:
- Clean the rubber door seal with a clean, damp, non-linting cloth;
- Check the chamber and shelves for cleanliness and debris;
- Fill the reservoir with distilled water or RO water;
- Turn the power source on.
If the steriliser fails to meet any of the test requirements, it should be withdrawn from service and advice should be sought from the manufacturer and/or maintenance contractor. Any instruments processed in an unsuccessful cycle should not be used.